Volume 32, Issue 7 (October 2021)                   Stud Med Sci 2021, 32(7): 558-571 | Back to browse issues page


XML Persian Abstract Print


Download citation:
BibTeX | RIS | EndNote | Medlars | ProCite | Reference Manager | RefWorks
Send citation to:

Zeinali A, Kargar N. COMPARISON OF THE PERFORMANCE OF MONTE-CARLO AND COLLAPSED CONE ALGORITHMS USED IN MONACO TREATMENT PLANNING SYSTEM IN PREDICTING CARDIOPULMONARY COMPLICATIONS DUE TO THE LEFT BREAST RADIOTHERAPY. Stud Med Sci. 2021; 32 (7) :558-571
URL: http://umj.umsu.ac.ir/article-1-5434-en.html
Department of Medical Physics, School of Medicine, Urmia University of Medical Sciences, Urmia, Iran , niloofar.jeddi@yahoo.com
Abstract:   (416 Views)
Background & Aims: The aim of this study was to evaluate the performance of dose calculation algorithms used in the Monaco treatment planning system to predict cardiopulmonary complications due to left breast radiotherapy.
Materials & Methods: Three-dimensional dose distribution of 21 patients with left breast cancer was prepared by two-dose calculation algorithms (CC and MC) with the same unit monitor. Cardiopulmonary complications due to radiation therapy in these patients were evaluated by different radiobiological models as well as various parameters extracted from previous studies using MATLAB software. In this study, MC dose calculation is considered as benchmark data. Algorithms, Friedman nonparametric test, and Wilcoxon test were used for statistical comparison of the obtained results.
Results: For both dose calculation algorithms, the value of TCP was estimated to be acceptable, with the same parameter being higher for the Poisson model than for the Niemierko model. The difference in NTCP for CC and MC algorithms for pulmonary pneumonitis, pericarditis, and cardiac mortality is not statistically significant by most parameters.
Conclusion: Both dose calculation algorithms estimate the TCP value as acceptable and the NTCP calculated by these two algorithms is close to the expected NTCP value. The value obtained for TCP, and NTCP depends on the radiobiological parameters used in the mathematical formula and the amount of dose extracted from the dose calculation algorithms.
Full-Text [PDF 1141 kb]   (182 Downloads)    
Type of Study: Research | Subject: فیزیک پزشکی

Add your comments about this article : Your username or Email:
CAPTCHA

Send email to the article author


Rights and permissions
Creative Commons License This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.